Determination of effective source to surface distance and cutout factor in small fields in electron beam radiotherapy: A comparison of different dosimeters

Tytuł:
Determination of effective source to surface distance and cutout factor in small fields in electron beam radiotherapy: A comparison of different dosimeters
Autorzy:
Bayatiani, Mohamad Reza
Fallahi, Fatemeh
Aliasgharzadeh, Akbar
Ghorbani, Mahdi
Khajetash, Benyamin
Seif, Fatemeh
Data publikacji:
2020
Słowa kluczowe:
radiotherapy
linac
electron beam
effective SSD
cutout
dosimetry
Język:
angielski
Dostawca treści:
BazTech
Artykuł
  Przejdź do źródła  Link otwiera się w nowym oknie
Objective: The main purpose of this study is to calculate the effective source to surface distance (SSDeff) of small and large electron fields in 10, 15, and 18 MeV energies, and to investigate the effect of SSD on the cutout factor for electron beams a linear accelerator. The accuracy of different dosimeters is also evaluated. Materials and methods: In the current study, Elekta Precise linear accelerator was used in electron beam energies of 10, 15, and 18 MeV. The measurements were performed in a PTW water phantom (model MP3-M). A Semiflex and Advanced Markus ionization chambers and a Diode E detector were used for dosimetry. SSDeff in 100, 105, 110, 115, and 120 cm SSDs for 1.5 × 1.5 cm2 to 5 × 5 cm2 (small fields) and 6 × 6 cm2 to 20 × 20 cm2 (large fields) field sizes were obtained. The cutout factor was measured for the small fields. Results: SSDeff in small fields is highly dependent on energy and field size and increases with increasing electron beam energy and field size. For large electron fields, with some exceptions for the 20 × 20 cm2 field, this quantity also increases with energy. The SSDeff was increased with increasing beam energy and field size for all three detectors. Conclusion: The SSDeff varies significantly for different field sizes or cutouts. It is recommended that SSDeff be determined for each electron beam size or cutout. Selecting an appropriate dosimetry system can have an effect in determining cutout factor.
Opracowanie rekordu ze środków MNiSW, umowa Nr 461252 w ramach programu "Społeczna odpowiedzialność nauki" - moduł: Popularyzacja nauki i promocja sportu (2021).

Ta witryna wykorzystuje pliki cookies do przechowywania informacji na Twoim komputerze. Pliki cookies stosujemy w celu świadczenia usług na najwyższym poziomie, w tym w sposób dostosowany do indywidualnych potrzeb. Korzystanie z witryny bez zmiany ustawień dotyczących cookies oznacza, że będą one zamieszczane w Twoim komputerze. W każdym momencie możesz dokonać zmiany ustawień dotyczących cookies