Informacja

Drogi użytkowniku, aplikacja do prawidłowego działania wymaga obsługi JavaScript. Proszę włącz obsługę JavaScript w Twojej przeglądarce.

Tytuł pozycji:

Mechanical extraction of implantable cardioverter-defibrillator leads with a dwell time of more than 10 years : insights from a single high-volume centre

Tytuł:
Mechanical extraction of implantable cardioverter-defibrillator leads with a dwell time of more than 10 years : insights from a single high-volume centre
Autorzy:
Boczar, Krzysztof
Lelakowski, Jacek
Holcman, Katarzyna
Pfitzner, Roman
Dębski, Maciej
Małecka, Barbara
Ząbek, Andrzej
Kostkiewicz, Magdalena
Musiał, Robert
Ulman, Mateusz
Data publikacji:
2023
Słowa kluczowe:
lead
cardiac implantable electronic devices
safety
complications
effectiveness
transvenous lead extraction
implantable cardioverter-defibrillator
Język:
angielski
ISBN, ISSN:
10995129
Prawa:
Udzielam licencji. Uznanie autorstwa - Użycie niekomercyjne 4.0 Międzynarodowa
http://creativecommons.org/licenses/by-nc/4.0/legalcode.pl
Linki:
https://academic.oup.com/europace/advance-article/doi/10.1093/europace/euac272/6993754  Link otwiera się w nowym oknie
Dostawca treści:
Repozytorium Uniwersytetu Jagiellońskiego
Artykuł
Aims. To analyze and compare the effectiveness and safety of transvenous lead extraction (TLE) of implantable cardioverter-defibrillator (ICD) leads with a dwell time of >10 years (Group A) vs. younger leads (Group B) using mechanical extraction systems. Methods and results. Between October 2011 and July 2022, we performed TLE in 318 patients. Forty-six (14.4%) extracted ICD leads in 46 (14.5%) patients that had been implanted for >10 years. The median dwell time of all extracted ICD leads was 5.9 years. Cardiovascular implantable electronic device-related infection was an indication for TLE in 31.8% of patients. Complete ICD leads removal and complete procedural success in both groups were similar (95.7% in Group A vs. 99.6% in Group B, P = 0.056% and 95.6% in Group A vs. 99.6% in Group B, P = 0.056, respectively). We did not find a significant difference between major and minor complication rates in both groups (6.5% in Group A vs. 1.5% in Group B and 2.2% in Group A vs. 1.8% in Group B, P = 0.082, respectively). One death associated with the TLE procedure was recorded in Group B. Conclusion. The TLE procedures involving the extraction of old ICD leads were effective and safe. The outcomes of ICD lead removal with a dwell time of >10 years did not differ significantly compared with younger ICD leads. However, extraction of older ICD leads required more frequent necessity for utilizing multiple extraction tools, more experience and versatility of the operator, and increased surgery costs.

Ta witryna wykorzystuje pliki cookies do przechowywania informacji na Twoim komputerze. Pliki cookies stosujemy w celu świadczenia usług na najwyższym poziomie, w tym w sposób dostosowany do indywidualnych potrzeb. Korzystanie z witryny bez zmiany ustawień dotyczących cookies oznacza, że będą one zamieszczane w Twoim komputerze. W każdym momencie możesz dokonać zmiany ustawień dotyczących cookies