Tytuł pozycji:
Electrolytic preconcentration of platinum for electrothermal atomic absorption spectrometry using a flow-through graphite tube cell
An electrodeposition method in a flow-through graphite tube cell coupled to an electrothermal atomic absorption spectrometer was developed for the preconcentration and determination of platinum. A flow system incorporating a three-electrode arrangement for this preconcentration was constructed. After the deposition step, the graphite tube was placed into the graphite furnace and an atomization program was applied. The utility of the proposed procedure was checked with the analysis of a series of samples in the range of platinum concentration from 0.1 to 10 ng ml (-1) for 30 min deposition.
Zastosowano technikę elektrolitycznego zatężania na elektrodzie grafitowej w układzie przepływowym do zatężania i oznaczania platyny metodą absorpcyjnej spektrometrii atomowej z atomizacją elektrotermiczną. Skonstruowano, w celu zatężania, system przepływowy składający się z układu trójelektrodowego. Po etapie zatężania, elektroda grafitowa była stosowana jako piec grafitowy w procesie atomizacji. Przydatność zaproponowanej metody sprawdzono analizując próbki zawierające platynę w zakresie od 0. l do 10 ng m(-1) stosując 30 min zatężanie.