On calibration of solid phase microextraction-gas chromatography-mass spectrometry system for analysis of organic air contaminants using gaseous standard mixtures
The paper briefly touches the general problems related to calibration in analytical measurements. The special requirements with respect to calibration in methods of air organic contaminants determination (those based on solid phase microextraction, SPME) are stressed. A convenient apparatus for generation of gaseous mixture with low organic analyte content with facilities for SPME sampling is described. The procedure for temperature and humidity corrected calibration is proposed and studies on effects of these parameters on caJibration curves are presented.
W pracy zwrócono uwagę na gJówne problemy związane z kalibracją w pomiarach analitycznych. Podkreślono specjalne wymagania w odniesieniu do kalibracji w przy-padku metod oznaczania organicznych zanieczyszczeń powietrza opartych na technice SoJid Phase Microextraction (SPME). Opisano wygodny aparat do generowania gazo-wych mieszanin wzorcowych o niskiej zawartości analitów organicznych, z którego w prosty sposób można pobierać zanieczyszczenia organiczne na włókno SPME. Zapro-ponowano procedurę kalibracji, w której uwzględnia się wpływ temperatury i wilgot-ności. Opisano także wpływ tych parametrów na krzywe kalibracyjne.