Tytuł pozycji:
Simultaneous determination of Ne, Ar, CFC-11, CFC-12 and SF6 in grounwater samples by gas chromatography
Ne, Ar, CFC-11, CFC-12 and SF6 are widely used as markers in the studies of ground-water systems. Usually, their analysis involves a use of two or three separate instrumental methods and separate sets of water samples. In this work a gas chromatographic method for simultaneous determination of Ne. Ar, SF6, CFC-11 and CFC-12 in one groundwater sample has been presented. The analytes of interest were extracted from the water sample using the headspace method and analyzed employing a system consisting of four chromatographic channels. Limits of detection for the measured tracers were: 0.06 fmol L-1 (0.0081 pg kg-1) for SF6,15 fmol L-1 (2.1 pg kgpg kg-1) for CFC-11, 10 fmol L-1 (1.23 pgkg-1) for CFC-12,0.84 nmol L-1 ( 1.9x 10 cm3 STP cm-3) for Ne, and 0.15 &mumol L-1 (3.1 x 10-6 cm-3 STP cm-3) for Ar.
Ne, Ar. CFC-11, CFC-12 i SSF6 sąpowszechnie stosowane jako znaczniki w badaniach wód podziemnych. Pomiar stężenia tych związków przeprowadza się zazwyczaj dwiema lub trzema różnymi metodami analitycznymi z wykorzystaniem oddzielnych zestawów próbek wód. W niniejszej pracy przedstawiono opis chromatograficznej metody jednoczesnego oznaczania Ne, Ar, SF6 oraz freonów F-11 i F-12 w pojedynczej próbce wód podziemnych.Poziomy wykrywalności mierzonych związków wynoszą 0,06 fmol L-1 (0.0081 pg kg w przypadku SF6; 15 fmol L-1 (2,1 fg cm-3 w przypadku F-ll; 10 fmol L<-1 (1,23 pg kg-1 w przypadku F-12; 0,84 nmol L-1 (l ,9 x 10-8 cm3 STP cm-') w przypadku Ne i 0,15 μmol L-1 (3,1 x 10-6 cm3 STP cm-3) w przypadku Ar.