Tytuł pozycji:
Determination of chromium(VI) and common inorganic anions in industrial wastewater by means of suppressed ion chromatography
Isocratic ion chromatography with suppressed conductivity detection was used for the simultaneous separation and determination of chloride, nitrate, sulfate and chromate in industrial wastewater. The separation was performed on an anion-exchange column with Na2CO,/NaHCO3 eluent and 20 Ś 10'* dm1 injection loop. Linearity of chromate was checked up to 30 mg/dm1. The detection limits of CT, NO3\ SO42~and CrO42" were on the levels: 1.18 mg/dm3, 0.31 mg/dm3, 1.74 mg/dm3, and 0.63 mg/dm3, respectively. The mean recoveries of target anions for spiked samples were 87-109% and accuracy did not exceed 4.67%.
Do jednoczesnego rozdzielania i oznaczania chlorków, azotanów, siarczanów i chromianów w ściekach przemysłowych zastosowano izokratyczną chromatografię jonową z detekcja konduktometryczną. Rozdzielanie przeprowadzono za pomocą kolumny anionowymiennej i eluentu Na2CO,/NaHCO, oraz pętli wstrzykowej o pojemności 20 o 10* dm1. Zakres liniowości dla chromianów sprawdzono od granic wykrywalności do 30 mg/dm'. Granice wykrywalności jonów Cl", NO,', SO42' i CrO42' były odpowiednio na poziomie: 1,18 mg/dm3, 0,31 mg/dm-1, 1,74 mg/dm1 i 0,63 mg/dm1. Średni odzysk oznaczanych anionów w wzbogacanych próbkach był w zakresie od 87% do 109%, a wartość RSD nie przekraczała 4,67%.