Point-of-care testing for C-reactive protein in a sequential microfluidic device

Tytuł:
Point-of-care testing for C-reactive protein in a sequential microfluidic device
Autorzy:
Szot-Karpińska, Katarzyna
Jönsson-Niedziółka, Martin
Niedziółka-Jönsson, Joanna
Boonkaew, Suchanat
Pałys, Barbara
Data publikacji:
2023
Wydawca:
Elsevier
Słowa kluczowe:
phage display
capillary-driven microfluidic device
microfluidics
C-reactive protein
electrochemical detection
point-of-care devices
peptides
label-free immunoassay
Język:
angielski
Prawa:
http://creativecommons.org/licenses/by/4.0/
Linki:
https://open.icm.edu.pl/handle/123456789/22975  Link otwiera się w nowym oknie
Dostawca treści:
Repozytorium Centrum Otwartej Nauki
Artykuł
  Przejdź do źródła  Link otwiera się w nowym oknie
This research was supported by the National Centre for Research and Development (NCBR) through the EEA and Norway Grants (Project number: NOR/POLNOR/UPTURN/0060/2019) and the Polish National Science Centre via a SONATA 13 grant UMO-2017/26/D/ST5/00980

Point-of-care testing (POCT) devices are currently available for the detection of various important diseases owing to their easy-to-use, portable, and user-friendly nature. However, the multistep regent manipulation still impedes the performance of the device for end-users. To address this issue, we present a novel sequential flow-through microfluidic device that incorporates dual flow behaviors (fast-flow/delayed) into a single device for assessing the risk of cardiovascular disease (CVDs) using C-reactive protein (CRP) as the model analyte. The device features a fast-flow channel for automated washing of unbound CRP antigens and a delayed channel for flow of a redox reagent for electrochemical analysis. It requires only a single buffer loading, resulting in a total analysis time of 15 min. To overcome challenges related to storage stability and the affinity of antibodies in the manufacturing process, we integrated the P3-CRP peptide as a new bioreceptor for CRP detection. The performance of both the P3-CRP peptide and antibodies was assessed and examined using chronocoulometry. The proposed device exhibited a limit of detection at 47 pg mL−1 with a wide linear range (5 orders of magnitude). To validate its applicability, the device was tested with serum, plasma, and whole blood samples, yielding satisfactory results.

Ta witryna wykorzystuje pliki cookies do przechowywania informacji na Twoim komputerze. Pliki cookies stosujemy w celu świadczenia usług na najwyższym poziomie, w tym w sposób dostosowany do indywidualnych potrzeb. Korzystanie z witryny bez zmiany ustawień dotyczących cookies oznacza, że będą one zamieszczane w Twoim komputerze. W każdym momencie możesz dokonać zmiany ustawień dotyczących cookies