Background: The objective of this work is to analyze the effectiveness of Penumbra 400 micro
-coils in the embolization of large cerebral aneurysms.
Material and methods: A retrospective analysis has been conducted in a group of 32 patients at
the average age of 54.5 years (30-84) for whose embolization the P400 micro-coils (P400) have
been used. A control group consisted of 44 patients at the average age of 52.7 years (24–82) in
whose aneurysm embolization the 18 micro-coils (MC) have been utilized.
Results: The respective percentages of micro-coil packing density in aneurysm sacs were
31.5% for P400 and 29% for MC. The average P400 fluoroscopy time was 21 min, and 34 min in
case of MC. The average number of used micro-coils was 3.9 for P400 and 5.6 for MC. The
radiation dose received by a patient was 1.7 Gy/2.2 Gy, respectively. The recanalization of
P400 has occurred in 14/31 cases (45%), and for MC it has occurred in 23/44 (52%) patients. One
patient died due to early recanalization after P400 aneurysm embolization.
Conclusion: Procedures with use of the P400 demonstrate minimally higher effectiveness of
large aneurysms embolization in comparison with the MS with a not much shorter duration
and reduction of a radiation dose that a patient receives.