Tytuł pozycji:
Improvement of copper FAAS determination conditions via preconcentration procedure with the use of salicylaldoxime complex trapped in polymer matrix
The paper presents application of a new resin dedicated to
copper(II) flow-injection on-line preconcentration prior its
flame atomic absorption spectrometric (FAAS) determination.
The new sorbent, obtained by suspension polymerization
technique, was styrene-divinylbenzene copolymer modified
with 5-dodecylsalicylaldoxime-copper(II) complex. In flow
mode leaching of initially imprinted Cu(II) ions from polymer
beads was effective with 1% (v/v) nitric acid, however for
elution of ions retained on the sorbent during the loading
process sufficient efficiency was obtained for 0.5% (v/v) nitric
acid. The most effective copper(II) sorption was observed
within sample pH ca. 6.3 at flow rate 7.5 mLmin$^{-1}$.
Furthermore, preconcentration studies of Cu(II) ions realized
in the presence of popular foreign ions like Cd(II), Pb(II), Zn(II), Ni(II), Mn(II), Co(II) did not reveal significant interference. The expected effect of Cu(II)-imprinting was confirmed by higher tolerance level for interferents ions
concentration for the new sorbent than for the control polymer.
It
was
found
that
alkaline
metals
ions
and
humic
acid
had
the
most
relevant
influence
on
copper(II)
uptake.
Accuracy
of
the
evaluated
method
was
assessed
for
anal-
ysis
of
water
samples
(tap
and
mineral
water,
river
water,
artesian
water)
and
certified
water
reference
materials
compare
them
to
results
obtained
by
inductively
plasma
mass
spectrometry. The satisfactory relative error values obtained with use of standard addition calibration method, confirms the feasibility of this method for Cu(II) determination in water samples. Application of 120s sorption time enabled to obtain 74-fold enrichment factor and limit of detection (3$\sigma$) equal to 0.4$\mu$gL$^{-1}$.