Comparative studies of the sensitivities of sparse and full geometries of Total-Body PET scanners built from crystals and plastic scintillators

Tytuł:
Comparative studies of the sensitivities of sparse and full geometries of Total-Body PET scanners built from crystals and plastic scintillators
Autorzy:
Stępień, Ewa
Krzemień, Wojciech
Kumar, Deepak
Perez del Rio, Elena
Wiślicki, W.
Dulski, Kamil
Moskal, Paweł
Curceanu, C.
Vandenberghe, S.
Klimaszewski, Konrad
Hiesmayr, B. C.
Shivani, Shivani
Chug, Neha
Gajos, Aleksander
Baran, Jakub
Dadgar, Meysam
Parzych, Szymon
Shopa, R. Y.
Kapłon, Łukasz
Panek, Dominik
Czerwiński, Eryk
Kozik, Tomasz
Korcyl, Grzegorz
Tayefi Ardebili, Faranak
Tayefi Ardebili, Keyvan
Valsan Eliyan, Kavya
Raczyński, L.
Niedźwiecki, Szymon
Sharma, Sushil
Konieczka, P.
Skurzok, Magdalena
Data publikacji:
2023
Słowa kluczowe:
GATE simulation
sensitivity
uEXPLORER
J-PET
Total-Body PET
Język:
angielski
ISBN, ISSN:
21977364
Prawa:
Udzielam licencji. Uznanie autorstwa 4.0 Międzynarodowa
http://creativecommons.org/licenses/by/4.0/legalcode.pl
Dostawca treści:
Repozytorium Uniwersytetu Jagiellońskiego
Artykuł
Background. Alongside the benefits of Total-Body imaging modalities, such as higher sensitivity, single-bed position, low dose imaging, etc., their final construction cost prevents worldwide utilization. The main aim of this study is to present a simulation-based comparison of the sensitivities of existing and currently developed tomographs to introduce a cost-efficient solution for constructing a Total-Body PET scanner based on plastic scintillators. Methods. For the case of this study, eight tomographs based on the uEXPLORER configuration with different scintillator materials (BGO, LYSO), axial field-of-view (97.4 cm and 194.8 cm), and detector configurations (full and sparse) were simulated. In addition, 8 J-PET scanners with different configurations, such as various axial field-of-view (200 cm and 250 cm), different cross sections of plastic scintillator, and multiple numbers of plastic scintillator layers (2, 3, and 4), based on J-PET technology have been simulated by GATE software. Furthermore, Siemens’ Biograph Vision has been simulated to compare the results with standard PET scans. Two types of simulations have been performed. The first one with a centrally located source with a diameter of 1 mm and a length of 250 cm, and the second one with the same source inside a water-filled cylindrical phantom with a diameter of 20 cm and a length of 183 cm. Results. With regards to sensitivity, among all the proposed scanners, the ones constructed with BGO crystals give the best performance ($\sim$ 350 cps/kBq at the center). The utilization of sparse geometry or LYSO crystals significantly lowers the achievable sensitivity of such systems. The J-PET design gives a similar sensitivity to the sparse LYSO crystal-based detectors while having full detector coverage over the body. Moreover, it provides uniform sensitivity over the body with additional gain on its sides and provides the possibility for high-quality brain imaging. Conclusion. Taking into account not only the sensitivity but also the price of Total-Body PET tomographs, which till now was one of the main obstacles in their widespread clinical availability, the J-PET tomography system based on plastic scintillators could be a cost-efficient alternative for Total-Body PET scanners.

Ta witryna wykorzystuje pliki cookies do przechowywania informacji na Twoim komputerze. Pliki cookies stosujemy w celu świadczenia usług na najwyższym poziomie, w tym w sposób dostosowany do indywidualnych potrzeb. Korzystanie z witryny bez zmiany ustawień dotyczących cookies oznacza, że będą one zamieszczane w Twoim komputerze. W każdym momencie możesz dokonać zmiany ustawień dotyczących cookies